l Four samples fused simultaneously, with auto-shaking and auto
control
l Auto control means the melting, shaking and standing steps are all
precisely repeatable
l Large screen display; auto-alarm and automatic power switching off
for overcurrent and overheat
l Customizable heating process, with several heating programs preset
Specs:
1. Heating unit: 3 pieces of
double helix ,single-end, silicon carbide rod
2. Maximum Temperature: 1250¡æ,continuously
adjustable
3. Temperature measurement unit: 2
pieces of platinum-rhodium thermocouple, with maximum measurable temperature
1600¡æ
4. Capacity: 4 samples
simultaneously (maximum 6 samples if required by user)
5. Wattage: 6kw
6. Power supply: 380-415V,50-60Hz£¬three phase
7. Dimension(Height ¡ÁWidth ¡Á
Depth): 500mm¡Á1217mm¡Á500mm
8. Weight: 115kg
Features:
1. Applies CPU microprocessor and
PID control technique, large screen Chinese display, and menu operation
2. Heating with double helix
silicon carbide rod; temperature measured with platinum-rhodium thermocouple
3. Up to 4 samples can be fused
simultaneously (maximum 6 samples if required by user)
4. Up to 10 customized heating
programs can be preset; within the programs, the heating process, the
temperature increase speed, the final temperature, the shaking time, shaking
amplitude, and shaking speed of the crucible stool, the cooling time, whether
to use fans, the air flow, and so on, all can be preset.
5. Automatic control in all the
process except for the filling of the fusion sample, enabling the sameness and
homogeneity of the sample; all samples are uniformly the same except for the
differences in composition
6. The bracket of the crucible and
the mould are made of high temperature alloy, which guarantees no chemical
reaction in high temperature
7. Chain wheel lever structure is
used in the opening device of the cover of the fusion room, trouble-free for
user operation
8. With minimum start temperature
alarm and power switch-off protection when the cover of the fusion room is open
9. Automatically switching off
power and setting alarm for overcurrent and overheat; independent temperature
measurement thermocouples are used for overheat protection and heating control;
together with the heating control circuit, they can form treble protection of
the system and can make the power switch-off protection for overheat more
reliable.